FDA Class IFDA·Medical devices·November 18, 2026

3B Medical recalls Luna G3 BPAP System over wrong pressure value being recorded and sent

When Targeted VT is enabled, the device records and transmits the configured IPAP value instead of the actual inspiratory pressure it delivers.

What to do

Check the device's model and its lot or serial number against this notice. If yours matches, contact the maker or your health care provider about what to do next, and don't stop using a device you depend on until you've talked to them.

Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.

Codes to check on your product

Look for these on the label, the packaging or the product itself. If they match, follow the steps above.

Codes
UDI-DI 06948538365731 Serial Numbers B3C22A17170 B3C22A17211 B3C22A17447 B3C22A18075 B3C22A18272 B3C22A18279 B3C22A18280 B3C22A18283 B3C22A18284 B3C22A18289 B3C22A18290 B3C22A18291 B3C22A18292 B3C22A18294 B3C22A18295 B3C22A18298 B3C22A18299 B3C22A18301 B3C22A18365 B3C22A18374 B3C22A18380 B3C22A18395 B3C22A18398 B3C22A18408 B3C22A18410 B3C22A18412 B3C22A18416 B3C22A18417 B3C22A18418 B3C22A18419 B3C22A18421 B3C22A18423 B3C22A18426 B3C22A18428 B3C22A18457 B3C22A20163 B3C22A21007 B3C22A21012 B3C22A21014 B3C22A21019 B3C22A21020 B3C22A21026 B3C22A21031 B3C22A21034 B3C22A21037 B3C22A21038 B3C22A21040 B3C22A21042 B3C22A21045 B3C22A21046 B3C22A21048 B3C22A21051 B3C22A21053 B3C22A21055 B3C22A21056…Usually printed or stamped on the label, the base or the cap.

What to do if something you own is recalled

Get an email when 3B Medical, Inc. has a new recall

Free, and no account needed. We’ll send a link to confirm your address, then an email only when a new recall appears.

Recall details

Product
Luna G3 BPAP System Model/Catalog Number: LG3800-30VT Product Description: The Luna G3 BPAP System is a Bi-level PAP (Bi-level Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA).
Hazard
The configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure when Targeted VT is enabled.
Units
12672
Where sold
US nationwide distribution including Puerto Rico. No OUS distribution.
FDA classification
Class I · Ongoing
Classified by FDA
November 12, 2026
Recall started
October 8, 2026
FDA enforcement report
November 18, 2026
Reference numbers
  • FDA recall number: Z-0409-2027

The headline, summary and steps above are written in plain English from the official notice. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.

Official notice

The notice's own wording from FDA.

Classification
Class I (Ongoing)
Quantity
12672