Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile…
A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active…
What to do
Check the device's model and its lot or serial number against this notice. If yours matches, contact the maker or your health care provider about what to do next, and don't stop using a device you depend on until you've talked to them.
Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.
Codes to check on your product
Look for these on the label, the packaging or the product itself. If they match, follow the steps above.
- Codes
- Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1Usually printed or stamped on the label, the base or the cap.
What to do if something you own is recalled
Get an email when Tandem Diabetes Care, Inc. has a new recall
Recall details
- Product
- Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app) Model/Catalog Number: 1015472
- Company
- Tandem Diabetes Care, Inc.
- Hazard
- A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.
- Units
- 4086 users
- Where sold
- Worldwide - US Nationwide and the country of Czech Republic.
- FDA classification
- Class II · Ongoing
- Classified by FDA
- November 12, 2026
- Recall started
- October 13, 2026
- FDA enforcement report
- November 18, 2026
- Reference numbers
- FDA recall number: Z-0437-2027
Details are taken from the official notice. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.
Official notice
The notice's own wording from FDA.
- Classification
- Class II (Ongoing)
- Quantity
- 4086 users
Related recalls
Tandem recalls t:slim X2 and Tandem Mobi insulin pumps over software defect that may cause unexpected insulin dose
A software defect in certain pump software versions, when paired with a G7 sensor, may cause an unexpected automatic insulin correction bolus (autobolus), which could result in hypoglycemia.
Tandem Diabetes Care, Inc. · 122,958 units
Tandem Diabetes Care recalls Tandem Mobi insulin pumps over false vibration motor failure from a software issue
A software issue can make the pump falsely report a vibration motor failure (Malfunction 12), so the pump cannot operate, the mobile app stops getting data, and insulin delivery and active CGM sessions stop, which could cause hyperglycemia.
Tandem Diabetes Care, Inc. · 17,745 units
Tandem recalls Tandem Mobi Insulin Pump over false blockage alarms that can pause insulin
Some Tandem Mobi insulin pumps can wrongly sound a blockage alarm when nothing is blocked, which suspends insulin delivery until the user acknowledges the alarm and resumes therapy, and repeated alarms may raise blood glucose.
Tandem Diabetes Care, Inc. · 80,938 units