title: "Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi… — Recalled Today"
description: "A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion…"
url: "https://recalledtoday.com/recall/fda-device-100091"
updated: "2026-11-25"
---

FDA·[Medical devices](https://recalledtoday.com/category/medical-devices)·November 18, 2026

# Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile…

A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active…

## What to do

Check the device's model and its lot or serial number against this notice. If yours matches, contact the maker or your health care provider about what to do next, and don't stop using a device you depend on until you've talked to them.

Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.

## Codes to check on your product

Look for these on the label, the packaging or the product itself. If they match, follow the steps above.

Codes

Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1Usually printed or stamped on the label, the base or the cap.

[What to do if something you own is recalled](https://recalledtoday.com/articles/what-to-do-if-your-product-is-recalled)

[FDA enforcement report](https://www.accessdata.fda.gov/scripts/ires/index.cfm?Event=100091)

## Get an email when Tandem Diabetes Care, Inc. has a new recall

Free, and no account needed. We’ll send a link to confirm your address, then an email only when a new recall appears.

## Recall details

Product

Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app) Model/Catalog Number: 1015472

Company

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care)

Hazard

A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.

Units

4086 users

Where sold

Worldwide - US Nationwide and the country of Czech Republic.

FDA classification

Class II · Ongoing

Classified by FDA

November 12, 2026

Recall started

October 13, 2026

FDA enforcement report

November 18, 2026

Reference numbers

- FDA recall number: Z-0437-2027

Details are taken from the official notice. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.

## Official notice

The notice's own wording from FDA.

Classification

Class II (Ongoing)

Quantity

4086 users

## Related recalls

FDA·[Medical devices](https://recalledtoday.com/category/medical-devices)

### [Tandem recalls t:slim X2 and Tandem Mobi insulin pumps over software defect that may cause unexpected insulin dose](https://recalledtoday.com/recall/fda-device-97403)

A software defect in certain pump software versions, when paired with a G7 sensor, may cause an unexpected automatic insulin correction bolus (autobolus), which could result in hypoglycemia.

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care) · 122,958 units

FDA Class IFDA·[Medical devices](https://recalledtoday.com/category/medical-devices)

### [Tandem Diabetes Care recalls Tandem Mobi insulin pumps over false vibration motor failure from a software issue](https://recalledtoday.com/recall/fda-device-97662)

A software issue can make the pump falsely report a vibration motor failure (Malfunction 12), so the pump cannot operate, the mobile app stops getting data, and insulin delivery and active CGM sessions stop, which could cause hyperglycemia.

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care) · 17,745 units

FDA Class IFDA·[Medical devices](https://recalledtoday.com/category/medical-devices)

### [Tandem recalls Tandem Mobi Insulin Pump over false blockage alarms that can pause insulin](https://recalledtoday.com/recall/fda-device-99851)

Some Tandem Mobi insulin pumps can wrongly sound a blockage alarm when nothing is blocked, which suspends insulin delivery until the user acknowledges the alarm and resumes therapy, and repeated alarms may raise blood glucose.

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care) · 80,938 units