title: "Tandem recalls Tandem Mobi Insulin Pump over false blockage alarms that can pause insulin — Recalled Today"
description: "Some Tandem Mobi insulin pumps can wrongly sound a blockage alarm when nothing is blocked, which suspends insulin delivery until the user acknowledges…"
url: "https://recalledtoday.com/recall/fda-device-99851"
updated: "2026-11-19"
---

FDA Class IFDA·[Medical devices](https://recalledtoday.com/category/medical-devices)·November 11, 2026

# Tandem recalls Tandem Mobi Insulin Pump over false blockage alarms that can pause insulin

Some Tandem Mobi insulin pumps can wrongly sound a blockage alarm when nothing is blocked, which suspends insulin delivery until the user acknowledges the alarm and resumes therapy, and repeated alarms may raise blood glucose.

## What to do

Check the device's model and its lot or serial number against this notice. If yours matches, contact the maker or your health care provider about what to do next, and don't stop using a device you depend on until you've talked to them.

Do not stop taking a prescribed medicine, or stop using a device you rely on, until you have talked to your pharmacist or doctor.

## Codes to check on your product

Look for these on the label, the packaging or the product itself. If they match, follow the steps above.

Codes

UDI-DI: 00389152075013 00389152271910 00389152350110 00389152365510 00389152365619 00389152370019 00389152480114 00389152481111 00389152481319. All affected pump models and serial numbers may be affected. Pump serial numbers with frequent false-positive occlusion alarms to be exchanged: 1198599 1199144 1201048 1201777 1202107 1202156 1204500 1204935 1207112 1209405 1209862 1211585 1218143 1220121 1220201 1220470 1220471 1221343 1221347 1221775 1221783 1222111 1223346 1223487 1223488 1223886 1223906 1223940 1224363 1224978 1225017 1225024 1225158 1225171 1225194 1225204 1225589 1225602 1225682 1226359 1226360 1226365 1226671 1226721 1226748 1226859 1226989 1227008 1227021 1227227 1227375…Usually printed or stamped on the label, the base or the cap.

[What to do if something you own is recalled](https://recalledtoday.com/articles/what-to-do-if-your-product-is-recalled)

[FDA enforcement report](https://www.accessdata.fda.gov/scripts/ires/index.cfm?Event=99851)

## Get an email when Tandem Diabetes Care, Inc. has a new recall

Free, and no account needed. We’ll send a link to confirm your address, then an email only when a new recall appears.

## Recall details

Product

Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078

Company

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care)

Hazard

Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm…

Units

80,938 (77,559 US, 3,379 OUS)

Where sold

Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada.

FDA classification

Class I · Ongoing

Classified by FDA

October 30, 2026

Recall started

September 23, 2026

FDA enforcement report

November 11, 2026

Reference numbers

- FDA recall number: Z-0316-2027

The headline, summary and steps above are written in plain English from the official notice. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.

## Official notice

The notice's own wording from FDA.

Classification

Class I (Ongoing)

## Related recalls

FDA Class IFDA·[Medical devices](https://recalledtoday.com/category/medical-devices)

### [Tandem Diabetes Care recalls Tandem Mobi insulin pumps over false vibration motor failure from a software issue](https://recalledtoday.com/recall/fda-device-97662)

A software issue can make the pump falsely report a vibration motor failure (Malfunction 12), so the pump cannot operate, the mobile app stops getting data, and insulin delivery and active CGM sessions stop, which could cause hyperglycemia.

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care) · 17,745 units

FDA·[Medical devices](https://recalledtoday.com/category/medical-devices)

### [Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile…](https://recalledtoday.com/recall/fda-device-100091)

A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals…

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care) · 4086 users

FDA·[Medical devices](https://recalledtoday.com/category/medical-devices)

### [Tandem recalls t:slim X2 and Tandem Mobi insulin pumps over software defect that may cause unexpected insulin dose](https://recalledtoday.com/recall/fda-device-97403)

A software defect in certain pump software versions, when paired with a G7 sensor, may cause an unexpected automatic insulin correction bolus (autobolus), which could result in hypoglycemia.

[Tandem Diabetes Care, Inc.](https://recalledtoday.com/company/tandem-diabetes-care) · 122,958 units