FDA·Healthcare and labs·September 9, 2026

Fresenius recalls 5008X dialysis blood tubing sets over risk of blood leak at a connection

The connection between the Arterial Alpha Clip and the main line or pump segment is more likely to leak, and a leak during treatment may cause blood loss and low blood volume (hypovolemia).

What to do

The notice does not give instructions. Read the official notice for details.

Recall details

Product
5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8; 5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6; 5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3
Hazard
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Units
6,624 units; 120 units; 62,544 units
Where sold
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
FDA classification
Class II · Ongoing
Classified by FDA
September 2, 2026
Recall started
August 10, 2026
FDA enforcement report
September 9, 2026
Identifiers
  • Codes: Lots 25JR01802 25NR01017 25NR01266 UDI-DI 00840861102525
  • FDA recall number: Z-3087-2026
  • Codes: Lots 25KR01227 25NR01257 UDI-DI 00840861102556
  • FDA recall number: Z-3084-2026
  • Codes: Lots 25JR01800 25KR01224 25LR01001 25NR01018 25NR01272 25PR01077 25PR01219 25PR01220 UDI-DI 00840861102518
  • FDA recall number: Z-3085-2026
  • Codes: Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532
  • FDA recall number: Z-3086-2026

The headline, summary and steps above are written in plain English from the official notice below. Identifiers are copied exactly as the agency published them. Always check the official notice before you act.

Official notice

The notice's own wording from FDA.

Product
5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8 | 5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6 | 5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3 | 5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C
Reason for recall
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Classification
Class II (Ongoing)
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Quantity
6,624 units; 120 units; 62,544 units; 18,600